Skip to content

Commit

Permalink
Separating pH from ORP
Browse files Browse the repository at this point in the history
Tested to work for pH (Y532) but not yet on ORP.
  • Loading branch information
aufdenkampe committed Jul 1, 2021
1 parent db97366 commit 4347622
Show file tree
Hide file tree
Showing 2 changed files with 30 additions and 4 deletions.
32 changes: 29 additions & 3 deletions src/YosemitechModbus.cpp
Original file line number Diff line number Diff line change
Expand Up @@ -205,7 +205,10 @@ bool yosemitech::startMeasurement(void)
if (respSize == 8 && modbus.responseBuffer[0] == _slaveID) return true;
else return false;
}
case Y532: // Does not require this function. Not described in the Y532 Modbus manual. Required to get Y532pH to work in ModularSensors.
// Y532 (pH) or Y533 (ORP) do not require Start/Stop functions. They are not listed in the Y532/Y533 Modbus Manual.
// However, Start/Stop functions are required to get these to work in ModularSensors.
case Y532:
case Y533:
{
return true;
}
Expand All @@ -230,6 +233,13 @@ bool yosemitech::stopMeasurement(void)
{
switch (_model)
{
// Y532 (pH) or Y533 (ORP) do not require Start/Stop functions. They are not listed in the Y532/Y533 Modbus Manual.
// However, Start/Stop functions are required to get these to work in ModularSensors.
case Y532:
case Y533:
{
return true;
}
case Y4000: // Does not require this function. Not described in the manual or sent using the MultiSensor_v1.18 software
{
return true;
Expand Down Expand Up @@ -290,20 +300,36 @@ bool yosemitech::getValues(float &parmValue, float &tempValue, float &thirdValue
}
break;
}
// Y532 (pH) or Y533 (ORP)
// Y532 (pH)
case Y532:
case Y533:
{
// According to the modbus manual we can get pH & potential starting at
// Register 0x2600, but it appears that the manual is not accurate.
if (modbus.getRegisters(0x03, 0x2800, 2))
{
parmValue = modbus.float32FromFrame(littleEndian, 3);
// Get temperature at register 0x2400
tempValue = modbus.float32FromRegister(0x03, 0x2400, littleEndian);
// Get potential (mV) at register 0x1200
thirdValue = modbus.float32FromRegister(0x03, 0x1200, littleEndian);
errorCode = 0x00; // No errors
return true;
}
break;
}
// //or Y533 (ORP)
// case Y533:
// {
// if (modbus.getRegisters(0x03, 0x2600, 4))
// {
// parmValue = modbus.float32FromFrame(littleEndian, 3);
// tempValue = modbus.float32FromRegister(0x03, 0x2400, littleEndian);
// thirdValue = modbus.float32FromRegister(0x03, 0x1200, littleEndian);
// errorCode = 0x00; // No errors
// return true;
// }
// break;
// }
// Y504 (DO)
case Y502:
case Y504:
Expand Down
2 changes: 1 addition & 1 deletion src/YosemitechModbus.h
Original file line number Diff line number Diff line change
Expand Up @@ -20,7 +20,7 @@ typedef enum yosemitechModel
Y516, // Oil in water?
Y520, // 4-Electrode Conductivity Sensor http://www.yosemitech.com/en/product-3.html
Y532, // pH
Y533, // ORP?
Y533, // ORP
Y550, // UV254 Sensor http://www.yosemitech.com/en/product-21.html
Y4000, // Multiparameter Sonde http://www.yosemitech.com/en/product-20.html
UNKNOWN // Use if the sensor model is unknown. Doing this is generally a
Expand Down